Title: | HIGH-SENSITIVITY CMUT WITH INTERLAYER METAL ARCHITECTURE FOR MEDICAL ULTRASONIC SYSTEMS |
Authors: | Cheng, Teng-Chuan Liao, Yu-Te Tsai, Tsung-Heng 電機學院 College of Electrical and Computer Engineering |
Keywords: | CMUT;TIA;ultrasonic sensor;transducer |
Issue Date: | 2015 |
Abstract: | This paper presents an integration of the capacitive micromachined ultrasonic transducers (CMUTs) and the transimpedance amplifier in medical ultrasound systems. The proposed CMUT and the sensing circuits are both fabricated on the same chip in a 0.18 mu m CMOS MEMS technology. Two interlayer metal structures are proposed to improve the sensitivity. The CMUTs are designed with 2MHz cells for medical diagnosis. The conversion efficiency of the CMUTs is 4 times and 1.8 times higher than that in the published structures. |
URI: | http://hdl.handle.net/11536/135764 |
ISBN: | 978-1-4799-8955-3 |
Journal: | 2015 TRANSDUCERS - 2015 18TH INTERNATIONAL CONFERENCE ON SOLID-STATE SENSORS, ACTUATORS AND MICROSYSTEMS (TRANSDUCERS) |
Begin Page: | 1211 |
End Page: | 1214 |
Appears in Collections: | Conferences Paper |